Brevis™ Transitional Implants

Description Angle Dia. Height Part Number
2.5 x 8.0mm 0˚ Brevis Transitional Implant 2.5mm 8.0mm 260-425-208
2.5 x 12.0mm 0˚ Brevis Transitional Implant 2.5mm 12mm 260-425-212

 

brev tran tranzitivni implantati

 

Brevis™ Transitional Implant Components

Description Part Number
Brevis™ Abutment Chairside Kit 260-100-212
Straight Brevis Latch Driver 260-101-902

 

o trans comp 1b tranzitivni implantati

 

brev latch driver tranzitivni implantati

 

 

Transitional Implant Considerations

  • Drill a 2.0mm osteotomy to appropriate depth with Bicon’s 2.0mm pilot drill with external irrigation.
  • Take care not to overheat the bone while drilling at ~1,100 RPM.
  • Pilot bur is marked at 6.0mm, 8.0mm, 11.0mm and 14.0mm.*
  • Be sure to drill to appropriate depth of intended implant to be placed or slightly beyond if the anatomy permits.
  • Make note of the quality of bone.
  • The Stealth and O-Ring implants are available in 8.0mm and 12.0mm lengths and either straight or angled.
  • Check angulation or trajectory of osteotomy with a paralleling pin.
  • Remove appropriate transitional implant from packaging and insert apical end of implant into osteotomy.
  • Assemble appropriate transitional implant driver to Bicon’s 400:1 handpiece or a comparable slow speed latch-type handpiece.
  • Slowly screw implant into the prepared osteotomy.

 

*Some pilot burs may have markings only at 8.0mm, 11.0mm, and 14.0mm. Check burs before drilling. If there is any doubt about the markings on any drill or reamer, take a measurement prior to using the reamer. (See Pilot Drill Reference).

 